Are CT Scans a Satisfactory Substitute for the Follow-Up of RSA Migration Studies of Uncemented Cups? A Comparison of RSA Double Examinations and CT Datasets of 46 Total Hip Arthroplasties

Table 1

Landmark errors in millimeters between RSA and CT data and between RSA double examinations after rigid body fitting. For the RSA double examinations markers with an error of the rigid body fitting above 0.35 mm are excluded automatically by the software. This is not the case for the comparison between RSA and CT.

VolumeMean errorMedianStandard deviationMinimumMaximum95% confidence interval

RSA-CT
Markers in bone0.270.220.170.021.500.00–0.61
Markers in prosthesis0.220.180.160.041.080.00–0.53
RSA1-RSA2 (double examination)
Markers in bone0.090.070.060.020.260.00–0.20
Markers in prosthesis0.110.090.060.020.260.00–0.24