Contrast Microsphere Destruction by a Continuous Flow Ventricular Assist Device: An In Vitro Evaluation Using a Mock Circulation Loop

Table 1

Prepump and postpump mean contrast signal intensity (dB) ± 1 SD with mean difference (run 1).

Prepump SI (dB)Postpump SI (dB)Mean difference

2400 RPM14.4 ± 0.910.3 ± 1.04.1 ± 1.3 < 0.0001
2800 RPM14.3 ± 0.811.1 ± 0.73.2 ± 1.1 < 0.0001
3200 RPM14.9 ± 0.612.1 ± 0.92.7 ± 1.2 < 0.0001
3600 RPM15.1 ± 0.612.8 ± 1.12.3 ± 1.3 < 0.0001