Research Article

Contrast Microsphere Destruction by a Continuous Flow Ventricular Assist Device: An In Vitro Evaluation Using a Mock Circulation Loop

Table 2

Prepump and postpump mean contrast signal intensity (dB) ± 1 SD with mean difference (run 2).

Prepump SI (dB)Postpump SI (dB)Mean difference

2400 RPM14.6 ± 0.79.7 ± 0.94.8 ± 1.6 < 0.0001
2800 RPM15.3 ± 0.711.1 ± 0.74.2 ± 1.1 < 0.0001
3200 RPM15 ± 0.712.2 ± 0.92.8 ± 1.2 < 0.0001
3600 RPM15.3 ± 0.613 ± 1.12.3 ± 1.3 < 0.0001