Contrast Microsphere Destruction by a Continuous Flow Ventricular Assist Device: An In Vitro Evaluation Using a Mock Circulation Loop
Table 3
Prepump and postpump mean contrast signal intensity (dB) ± 1 SD with mean difference (combined data).
| | Prepump SI (dB) | Postpump SI (dB) | Mean difference | |
| 2400 RPM | 14.5 ± 0.7 | 10.0 ± 1.0 | 4.4 ± 1.2 | < 0.0001 | 2800 RPM | 14.8 ± 0.9 | 11.1 ± 0.7 | 3.7 ± 1.2 | < 0.0001 | 3200 RPM | 14.9 ± 0.6 | 12.2 ± 0.9 | 2.8 ± 1.2 | < 0.0001 | 3600 RPM | 15.1 ± 0.6 | 12.9 ± 1.1 | 2.3 ± 1.3 | < 0.0001 |
|
|