Analytical Validation of a New Enzymatic and Automatable Method for d-Xylose Measurement in Human Urine Samples
Table 2
Results of the cross-validation experiment in each automated analyzer.
| | Cobas c502 (Roche) | ILab 600 (Werfen) | Dimension Vista 1500 (Siemens) |
| Urine LL samples (0.55 mg/dL) | | | | Mean (mg/dL) | 0.730 | 0.644 | 0.522 | SD | 0.006 | 0.013 | 0.010 | CV (%) | 0.8 | 2.0 | 1.8 | Inaccuracy (%) | 32.7 | 17.2 | 5.1 | Urine LQ samples (1.50 mg/dL) | | | | Mean (mg/dL) | 1.680 | 1.590 | 1.356 | SD | 0.012 | 0.013 | 0.184 | CV (%) | 0.7 | 0.8 | 13.5 | Inaccuracy (%) | 12.0 | 6.0 | 9.6 | Urine MQ samples (6.00 mg/dL) | | | | Mean (mg/dL) | 6.270 | 6.776 | 6.032 | SD | 0.021 | 0.037 | 0.054 | CV (%) | 0.3 | 0.6 | 0.9 | Inaccuracy (%) | 4.5 | 12.9 | 0.5 | Urine HQ samples (11.50 mg/dL) | | | | Mean (mg/dL) | 11.880 | 11.748 | 12.022 | SD | 0.025 | 0.039 | 0.122 | CV (%) | 0.2 | 0.3 | 1.0 | Inaccuracy (%) | 3.3 | 2.2 | 4.5 |
|
|