Research Article

Comparison of Complication and Success Rates of ProGlide Closure Device in Patients Undergoing TAVI and Endovascular Aneurysm Repair

Table 3

Comparison of results for PP closure device applied in patients undergoing TAVI and EVAR.

VariablesTAVI (N:58)EVAR (N:16)p

Complication (n,)2 (3.4)3 (18.8)0.031

Device failure (n,)2 (3.4)3 (18.8)0.031

Number of PP closure deviceused2.3±0.64.5±1.4<0.001

Right CFA diameter (mm)8.6±1.57.7±1.40.186

Left CFA diameter (mm)8.8±1.57.8±1.20.172

CFA: common femoral artery, EVAR: endovascular aortic aneurysm repair, and TAVI: transcatheter aortic valve implantation.