Research Article

Biomechanical Evaluation of a Novel Apatite-Wollastonite Ceramic Cage Design for Lumbar Interbody Fusion: A Finite Element Model Study

Table 2

Maximum stress values of the whole construct and their localisations under different loads. The maximum stress values of the whole construct by using K2 and K4 designs were significantly better (). I: intact model.

ā€‰FlexionExtensionBendingRotation
Max. stress valueMax. stress pointMax. stress valueMax. stress pointMax. stress valueMax. stress pointMax. stress valueMax. stress point

I93.5Disc
space
101.4Posterior
elements
85.2Disc
space
66.0Disc
space
K1112.1Disc
space
61.6Disc
space
100.8Disc
space
72.3Disc
space
K248.8Posterior
elements
33.9Posterior
elements
46.0Posterior
elements
63.6Posterior
elements
K3114.6Disc
space
68.4Disc
space
112.1Disc
space
80.6Disc
space
K448.7Posterior
elements
33.6Posterior
elements
48.6Disc
space
63.2Posterior
elements