A Simple Scatter Reduction Method in Cone-Beam Computed Tomography for Dental and Maxillofacial Applications Based on Monte Carlo Simulation

Table 2

Comparison of contrast in the selected regions before and after scatter reduction on the reconstructed images of real patients.

Patient caseReconstructed imagesContrast (grayscale value)
Air boneBone tissue

Case 1Before scatter reduction384.35261.97
After scatter reduction473.62333.42
Absolute percentage change23.22%27.27%

Case 2Before scatter reduction405.72262.71
After scatter reduction471.49309.16
Absolute percentage change16.21%17.68%

Case 3Before scatter reduction378.94223.56
After scatter reduction465.01280.51
Absolute percentage change22.71%25.47%

Case 4Before scatter reduction393.35238.34
After scatter reduction458.49287.16
Absolute percentage change16.56%20.48%