A Simple Scatter Reduction Method in Cone-Beam Computed Tomography for Dental and Maxillofacial Applications Based on Monte Carlo Simulation
Table 2
Comparison of contrast in the selected regions before and after scatter reduction on the reconstructed images of real patients.
| Patient case | Reconstructed images | Contrast (grayscale value) | Air bone | Bone tissue |
| Case 1 | Before scatter reduction | 384.35 | 261.97 | After scatter reduction | 473.62 | 333.42 | Absolute percentage change | 23.22% | 27.27% |
| Case 2 | Before scatter reduction | 405.72 | 262.71 | After scatter reduction | 471.49 | 309.16 | Absolute percentage change | 16.21% | 17.68% |
| Case 3 | Before scatter reduction | 378.94 | 223.56 | After scatter reduction | 465.01 | 280.51 | Absolute percentage change | 22.71% | 25.47% |
| Case 4 | Before scatter reduction | 393.35 | 238.34 | After scatter reduction | 458.49 | 287.16 | Absolute percentage change | 16.56% | 20.48% |
|
|