Journals
Publish with us
Publishing partnerships
About us
Blog
BioMed Research International
Journal overview
For authors
For reviewers
For editors
Table of Contents
Special Issues
BioMed Research International
/
2018
/
Article
/
Tab 1
/
Research Article
Increase in the Tibial Slope in Unicondylar Knee Replacement: Analysis of the Effect on the Kinematics and Ligaments in a Weight-Bearing Finite Element Model
Table 1
Overview of the material properties used in the FE simulation.
Deformable bodies (linear elastic)
Structure
Young’s modulus (MPa)
Poisson ratio
Structure
Young’s modulus (MPa)
Poisson ratio
Femoral component
210,000
0.3
Patella Cartilage
5.0
0.4
Inlay (UHMWPE)
312.5
0.46
Tibial Component
210,000
0.3
Ligaments (stiffness)
Structure
Initial strain
Stiffness (N/mm)
Structure
Initial strain
Stiffness (N/mm)
PCLa
-0.10
31.25
MCLa
0.02
27.9
PCLp
-0.02
15.0
MCLo
0.02
21.1
LCL
0.02
91.3
MCLs
0.02
72.2
ACLa
0.02
108.0
ACLp
0.02
108.0