Clinical Study
Comparative Evaluation of a New Depth of Anesthesia Index in ConView® System and the Bispectral Index during Total Intravenous Anesthesia: A Multicenter Clinical Trial
Table 4
The values of Ai and BIS during LOC.
| ||||||||||||||||||||||||||||||||||||
Data are presented as mean ± SD. Ai changed far more obviously than BIS from LOC to one minute after LOC (-9.13±10.20 VS -5.83±9.63, p<0.001). |