Research Article

The RITMIA™ Smartphone App for Automated Detection of Atrial Fibrillation: Accuracy in Consecutive Patients Undergoing Elective Electrical Cardioversion

Table 2

Ritmia Automated Diagnosis Compared To 12-Lead ECG Diagnosis.

12-Lead ECGAPP AFAPP NO-AFTotal

AF96399
Flutter088
SR47983
Total10090TOT=190