Comparisons of Effective Fields of Two Ultra-Widefield Ophthalmoscopes, Optos 200Tx and Clarus 500

Table 3

Comparison of appearance probability between “O > C” and “O < C” at four retinal quadrants by two raters.

Appearance probability of “O > C” (95% C.I.)Appearance probability of “O < C” (95% C.I.) value

Upper temporalRater 175.9 (62.4–86.5)24.1 (13.5–37.6)<0.001
Rater 272.9 (58.2–84.7)27.1 (15.3–41.8)0.002

Lower temporalRater 155.1 (40.2–69.3)44.9 (30.7–59.8)0.568
Rater 253.7 (37.4–69.3)46.3 (30.7–62.6)0.755

Upper nasalRater 152.4 (36.4–68.0)47.6 (32.0–63.6)0.878
Rater 257.1 (39.4–73.7)42.9 (26.3–60.6)0.500

Lower nasalRater 126.7 (14.6–41.9)73.3 (58.1–85.4)0.002
Rater 228.2 (15.0–44.9)71.8 (55.1–85.0)0.009

Statistical comparisons were performed by binomial tests. The results were considered statistically significant when . C.I., confidential interval.