Comparisons of Effective Fields of Two Ultra-Widefield Ophthalmoscopes, Optos 200Tx and Clarus 500
Table 3
Comparison of appearance probability between “O > C” and “O < C” at four retinal quadrants by two raters.
| | | Appearance probability of “O > C” (95% C.I.) | Appearance probability of “O < C” (95% C.I.) | value |
| Upper temporal | Rater 1 | 75.9 (62.4–86.5) | 24.1 (13.5–37.6) | <0.001 | Rater 2 | 72.9 (58.2–84.7) | 27.1 (15.3–41.8) | 0.002 |
| Lower temporal | Rater 1 | 55.1 (40.2–69.3) | 44.9 (30.7–59.8) | 0.568 | Rater 2 | 53.7 (37.4–69.3) | 46.3 (30.7–62.6) | 0.755 |
| Upper nasal | Rater 1 | 52.4 (36.4–68.0) | 47.6 (32.0–63.6) | 0.878 | Rater 2 | 57.1 (39.4–73.7) | 42.9 (26.3–60.6) | 0.500 |
| Lower nasal | Rater 1 | 26.7 (14.6–41.9) | 73.3 (58.1–85.4) | 0.002 | Rater 2 | 28.2 (15.0–44.9) | 71.8 (55.1–85.0) | 0.009 |
|
|
Statistical comparisons were performed by binomial tests. The results were considered statistically significant when . C.I., confidential interval. |