Clinical Study
The Accuracy of Computer-Assisted Implant Surgery Performed Using Fully Guided Templates versus Pilot-Drill Guided Templates
Table 4
Deviation values between planned and placed implants in the 2 study groups.
| | Fully drill guided | Pilot drill guided | | | Mean ± SD | Range (min-max) | Mean ± SD | Range (min-max) | P-value |
| Coronal Linear deviation (mm) | 1.16 ± 0.68 | 0.19 - 2.92 | 1.11 ± 1.05
| 0.20 - 5.04
| P=0.35 |
| Apical Linear Deviation (mm) | 1.65 ± 1.17 | 0.0 - 4.75 | 1.70 ± 1.12
| 0.0 - 4.31
| P=0.22 |
| Depth Linear Deviation (mm) | 0.95 ± 1.70 | 0.0 - 5.42 | -0.68 ± 1.31
| 0.0 - 4.11
| P=0.032 |
| Buccolingual Angular deviation (°) | 4.16 ± 3.59 | 0.0 - 12.05 | 6.73 ± 4.70
| 0.0 - 20.37
| P=0.042 |
| Mesiodistal Angular deviation (°)
| 2.81 ± 3.89 | 0.0 - 13.02 | 5.62 ± 5.19 | 0.0 - 22.25 | P=0.029 |
|
|