The maximum stress of Von Mises at 30 degrees.

Maximum equivalent stress (MPa)Static resistance of titanium alloy (MPa)%
In-Kone® Universal936.93102091.8%

Maximum equivalent tension value on bone tissue (MPa)Static bone resistance (MPa)%
In-Kone® Universal162.5218090.3%