Research Article

Comparison of Cochlear Implant Magnets and Their MRI Artifact Size

Table 1

Individual measurements at 1.5 T.

Distance of measurementABMedelOticon
IpsilateralContralateralIpsilateralContralateralIpsilateralContralateral

Area of measurementin cm
Volunteer 3Hard artifact3.810.93.210.44.410.7
Soft artifact4.7103.99.84.810
Survey sagittal3.73.63.7
Survey coronal6.166
Volunteer 2Hard artifact3.210.3310.80.310.3
Soft artifact58.54.78.64.19.3
Survey sagittal3.63.53.6
Survey coronal6.36.15.5

The table shows each measurement for each brand in terms of estimation (ipsilateral and contralateral) and the differentiation between soft and hard . In addition, the survey calculation is added