Research Article
Evaluation of Loopamp Assay for the Diagnosis of Pulmonary Tuberculosis in Cambodia
Table 2
Sensitivity and specificity of LED-fluorescent microscopy, TB-LAMP, and Xpert MTB/RIF.
| Test | Culture result | Sensitivity among culture-positive MTB | Specificity % (95% CI) | MTB-positive () positive/negative | MTB-negative () positive/negative | Smear-positive % (95% CI) | Smear-negative % (95% CI) | Overall % (95% CI) |
| Smear microscopy center | | | | | | | Microscopy | 109/48 | 6/336 | — | — | 69.4 (62.2-76.6) | 98.2 (96.8-99.6) | TB-LAMP | 128/29 | 9/333 | 94.5 (90.2-98.8) | 52.1 (38.0-66.2) | 81.5 (74.5-87.6) | 97.4 (95.7-99.1) | Xpert | 150/7 | 11/331 | 99.1 (93.6-99.9) | 87.5 (74.2-94.4) | 95.5 (92.3-98.8) | 96.8 (94.9-98.7) | Central laboratory | | | | | | | TB-LAMP | 90/7 | 4/95 | 100 | 56.3 (29.8-79.6) | 92.8 (87.6-97.9) | 96.0 (92.1-99.8) | Xpert | 88/9 | 3/96 | 100 | 43.8 (20.4-70.2) | 90.7 (85.0-96.5) | 97.0 (93.6-1.0) |
|
|