Beneficial Effects of Pulsed Electromagnetic Field during Cast Immobilization in Patients with Distal Radius Fracture

Table 2

Differences in the evaluation of pain, circumference, and range of motion between the PEMF and control groups during the study.

PEMF group (n = 27)Control group (n = 25) between groups

Pain VAS (mm)Before6.2 ± 2.007.1 ± 1.880.0956
After 3 weeks2.1 ± 1.413.5 ± 1.710.0052
value I vs. II
After 6 weeks0.2 ± 0.422.7 ± 2.15<0.0001
value I vs. III

Circumference of forearm (cm)Before25.8 ± 2.4625.9 ± 1.890.9198
After 3 weeks24.7 ± 2.3924.2 ± 1.940.5097
value I vs. II
After 6 weeks23.9 ± 2.3122.9 ± 1.950.1285
value I vs. III

Circumference of shoulder (cm)Before28.1 ± 2.5428.9 ± 3.510.4582
After 3 weeks26.9 ± 2.5227.6 ± 3.340.5097
value I vs. II
After 6 weeks26.4 ± 2.3626.8 ± 3.410.7486
value I vs. III

Range of motion (°) shoulder flexionBefore151.8 ± 18.96146.4 ± 17.320.1331
After 3 weeks162.1 ± 12.78156.6 ± 13.870.0280
value I vs. II
After 6 weeks167.5 ± 6.42162.0 ± 9.430.0034
value I vs. III

Shoulder extensionBefore39.6 ± 5.2337.8 ± 6.030.1935
After 3 weeks47.7 ± 3.1142.6 ± 5.190.0004
value I vs. II
After 6 weeks49.7 ± 0.8145.8 ± 4.290.0004
value I vs. III

Shoulder abductionBefore149.9 ± 18.21142.3 ± 14.120.0727
After 3 weeks164.5 ± 8.19153.6 ± 12.70.0015
value I vs. II
After 6 weeks168.4 ± 3.72159.5 ± 10.010.0002
value I vs. III

Elbow flexionBefore116.1 ± 14.10111.2 ± 8.550.2198
After 3 weeks129.8 ± 14.10122.8 ± 7.430.0115
value I vs. II
After 6 weeks141.6 ± 7.19132.1 ± 8.240.0001
value I vs. III

PEMF: pulsed electromagnetic fields; VAS : Visual Analogue Scale.