Clinical Study

Clinical Performance of the Automated LIAISON® Meridian H. pylori SA Stool Antigen Test

Table 2

LIAISON® Meridian H. pylori SA assay clinical performance in relation to the Composite Reference Method (CRM).

LIAISON® Meridian H. pylori SA
95% CI

Sensitivity95.5%87.5-99.1
Specificity97.6%94.5-99.2
Positive predictive value92.8%84.3-96.8
Negative predictive value98.6%95.8-99.5
Interrater agreement (kappa)0.9220.870-0.975