Clinical Study

Clinical Performance of the Automated LIAISON® Meridian H. pylori SA Stool Antigen Test

Table 4

Reproducibility of the LIAISON® Meridian H. pylori SA test assessed over 5 days at multiple sites.

Mean index valueWithin runDay to day within siteSite to siteTotal
SD%CVSD%CVSD%CVSD%CV

Neg Ctrl0.070.0045.10%0.0022.10%0.00912.50%0.01013.90%
Pos Ctrl4.800.0761.60%0.0631.30%0.1052.20%0.1533.10%
Mod Pos sample #12.120.0341.60%0.1085.10%0.1195.60%0.1688.00%
Mod Pos sample #22.370.0492.10%0.1566.60%0.2269.50%0.28311.90%
High Neg sample #10.690.0243.50%0.0375.40%0.0659.40%0.08111.80%
High Neg sample #20.690.0233.30%0.0192.70%0.0659.40%0.07711.00%
Low Pos sample #11.210.0312.50%0.0292.40%0.0937.70%0.1099.00%
Low Pos sample #21.200.0211.70%0.0564.70%0.12010.10%0.13811.50%