Research Article
An In Vitro Evaluation of the Red Cell Damage and Hemocompatibility of Different Central Venous Catheters
Table 5
Measurements from single-lumen CVCs after 500 mL/min flow rate of erythrocyte concentrate.
| | A () | B () | C () | Mean | SE | Mean | SE | Mean | SE |
| Total Hb (g/L) | 216.5 | 0.85 | 198.45 | 3.71 | 222.86 | 1.51 | Hematocrit (%) | 0.75 | 0 | 0.7 | 0.01 | 0.78 | 0.01 | Microparticles (MP/mL) | 268,723 | 54,641 | 434,352 | 83,019 | 257,923 | 35,026 | Erythrocyte MP (MP/mL) | 110,036 | 21,764 | 173,025 | 33,484 | 111,177 | 18,920 | Medium speed (cm/s) | 91.84 | 0 | 117.99 | 0 | 91.84 | 0 | Shear rate (1/s) | 540.24 | 0 | 786.59 | 0 | 540.24 | 0 | Shear stress (Pa) | 2.33 | 0.008 | 3.24 | 0.038 | 2.39 | 0.011 | Exposure time (s) | 0.6 | 0 | 0.6 | 0 | 0.6 | 0 | Dynamic viscosity (Pas) | 0.00432 | 0.000014 | 0.00412 | 0.000048 | 0.00443 | 0.000021 |
|
|