Application of 3D-Printed Orthopedic Cast for the Treatment of Forearm Fractures: Finite Element Analysis and Comparative Clinical Assessment
Table 5
Finite element analysis results.
| Loading condition | Displacement (mm) | Stress (MPa) | Global | Cast | Bone | Fracture surface | Cast | Bone | Fracture surface |
| AP | 12.685 | 11.959 | 6.824 | 1.265 | 59.375 | 174.720 | 4.599 | PA | 6.655 | 6.168 | 5.887 | 1.325 | 41.730 | 146.780 | 3.913 | ML | 3.437 | 3.123 | 2.706 | 0.850 | 22.607 | 86.248 | 5.033 | LM | 3.536 | 3.427 | 2.686 | 0.896 | 20.389 | 81.868 | 7.920 | IR | 0.226 | 0.226 | 0.166 | 0.026 | 2.603 | 6.067 | 0.173 | OR | 0.226 | 0.226 | 0.166 | 0.026 | 2.603 | 6.067 | 0.172 |
|
|
Note: data were generated by software calculation (no standard deviation).
|