Application of 3D-Printed Orthopedic Cast for the Treatment of Forearm Fractures: Finite Element Analysis and Comparative Clinical Assessment

Table 5

Finite element analysis results.

Loading conditionDisplacement (mm)Stress (MPa)
GlobalCastBoneFracture surfaceCastBoneFracture surface

AP12.68511.9596.8241.26559.375174.7204.599
PA6.6556.1685.8871.32541.730146.7803.913
ML3.4373.1232.7060.85022.60786.2485.033
LM3.5363.4272.6860.89620.38981.8687.920
IR0.2260.2260.1660.0262.6036.0670.173
OR0.2260.2260.1660.0262.6036.0670.172

Note: data were generated by software calculation (no standard deviation).