Table 4

Comparison of fit and accuracy of RPD frameworks constructed with conventional and digital techniques.

Digital frameworkMaterialFit accuracy (μm)
MeanSD

CAD/CAMCo-Cr alloy and PEEK41.67715.546
SLM/SLSCo-Cr alloy97.45232.575
Conventional framework
Lost wax techniqueCo-Cr alloy114.06377.704

Co-Cr: cobalt chromium; PEEK: polyetheretherketone; CAD/CAM: computer-aided designing and computer-aided manufacturing; SLM: selective laser melting; SD: standard deviation; SLS: selective laser sintering; μm: micrometer; RPD: removable partial denture.