The Study of Field Equivalence Determined by the Modeled Percentage Depth Dose in Electron Beam Radiation Therapy

Table 4

The comparison ofand, the minimum side lengths to remain the standard percent depth doses of 9 MeV and 15 MeV unchanged are 4.9 cm and 5.9 cm, respectively.

Energy (MeV) (nominal) (MeV)Slope (80%-20%)tan-1 (slope) (mm, this study) (mm)

65.41903.1-8.6593-1.455820.001-6.907738.1336.75
98.662106.2-5.7176-1.397650.0007-7.264449.1746.50
1211.272908.9-4.5559-1.354730.0005-7.600953.9853.04
1514.0440012.3-3.3749-1.282730.0003-8.111759.3359.20
1817.0347015.1-2.6474-1.209640.0002-8.517162.1065.20