Detailed results of microorganisms that either were misidentified or could not be identified by one or both MALDI-TOF MS systems.

Test groupsReferenceAutof ms1000Bruker Biotyper
ResultRangeAgreementResultScoreAgreement

Preserved strainsStreptococcus pneumoniaeS. pneumoniae9.400ID speciesS. pneumoniae1.924ID genus
Serratia nematodiphilaS. marcescens9.431MisID speciesS. ureilytica2.136MisID species
Candida orthopsilosisC. orthopsilosis9.507ID speciesC. orthopsilosis1.82ID genus
Streptococcus parasanguinisS. parasanguinis9.361ID speciesS. parasanguinis1.907ID genus
Candida glabrataC. glabrata9.441ID speciesC. glabrata1.866ID genus
Eggerthella lentaE. lenta8.967ID genusE. lenta2.114ID species
Acinetobacter bereziniaeA. guillouiae9.127MisID speciesA. bereziniae2.177ID species
Candida aurisNo reliable ID5.401No IDC. auris2.031ID species
Candida aurisNo reliable ID4.755No IDC. auris1.892ID genus
Clinical isolatesCandida aurisNo reliable ID4.952No IDC. auris2.073ID species
Enterococcus gallinarumE. casseliflavus9.543MisID speciesE. casseliflavus1.905ID genus
Klebsiella oxytocaRaoultella ornithinolytica9.159MisID speciesK. oxytoca2.173ID species
Staphylococcus epidermidisS. epidermidis9.508ID speciesS. epidermidis1.982ID genus