Research Article
Effect of Calcar Screw in Locking Compression Plate System for Osteoporotic Proximal Humerus Fracture: A Finite Element Analysis Study
Table 3
Comparison of maximum stresses applied to each primary component with respect to fracture models and load conditions. LCP: locking compression plate.
| Load condition | Fracture model | Max. shear stress on cortical bone (MPa) | Max. von-Mises stress on LCP (MPa) | Max. von-Mises stress on screw (MPa) | Type | Magnitude |
| Axial force | 500 N | A | 50.80 | 20.54 | 42.78 | B | 53.96 | 20.69 | 42.47 | C | 118.5 | 411.32 | 351.84 | D | 261.92 | 1550.30 | 913.20 | Shear force | 500 N/20° | A | 65.05 | 101.15 | 155.08 | B | 64.77 | 109.15 | 143.04 | C | 251.10 | 788.66 | 663.40 | D | 687.86 | 2926.80 | 1277.50 |
|
|