Comparison of maximum stresses applied to each primary component with respect to fracture models and load conditions. LCP: locking compression plate.

Load conditionFracture modelMax. shear stress on cortical bone (MPa)Max. von-Mises stress on LCP (MPa)Max. von-Mises stress on screw (MPa)
TypeMagnitude

Axial force500 NA50.8020.5442.78
B53.9620.6942.47
C118.5411.32351.84
D261.921550.30913.20
Shear force500 N/20°A65.05101.15155.08
B64.77109.15143.04
C251.10788.66663.40
D687.862926.801277.50