Review Article
Advances in Cuffless Continuous Blood Pressure Monitoring Technology Based on PPG Signals
Table 3
Summary of BP estimation methods based on PTT/PAT/PWV.
| Authors | Signals | Position of sensors | Subjects | Results (mmHg) | SBP | DBP |
| Viunytskyi et al. [37] | PPG, ECG | Finger, chest | 30 records | RMSE: 5.71 | RMSE: 5.13 | Lazazzera et al. [38] | PPG, PPG | Wrist, finger | subjects | ME: -1.52 STD: 9.45 | ME: 0.39 STD: 4.93 | Kim et al. [39] | PPG, PPG | Finger, finger | 21 subjects | | Byfield et al. [40] | PPG, PPG | Finger, finger | 26 subjects | MAE: 2.117 STD: 0.257 | MAE: 2.935 STD: 0.721 | Tabei et al. [41] | PPG, PPG | Finger, finger | 6 subjects | MAE: 2.07 STD: 2.06 | MAE: 2.12 STD: 1.85 | Marzorati et al. [42] | PPG, PCG | Finger, chest | 20 subjects | MAE: 3.06 | MAE: 1.83 | Huynh et al. [43] | PPG, IPG | Finger, wrist | 15 subjects | RMSE: 8.47 STD: 0.91 | RMSE: 5.02 STD: 0.73 | Yousefian et al. [44] | PCD, BCG | Wrist, wrist | 22 subjects | MAE: 7.6 | MAE: 5.1 |
|
|