Research Article

Fabrication and Optimization of ChE/ChO/HRP-AuNPs/c-MWCNTs Based Silver Electrode for Determining Total Cholesterol in Serum

Table 6

Comparison of the present method with previously reported biosensor for total cholesterol determination.

TransducerMethod of enzyme immobilizationWorking potentialResponse timeDetection limitLinearityStorage stabilityReference

Laponite clay nanoparticles-pol((12-pyrrol-1-dodecyl)triethylammonium tetrafluoroborate)/Pt disk electrodeChO, ChE enzyme Entrapment0.53 V versus Ag/AgCl50 sec20 M20 days[39]

Screen printed graphite electrodeChO, ChE, HRP, K4Fe(CN)6
Physical adsorption
−0.2 V versus Ag/AgCl2.81 mM2.81–13 mM[40]

Polydiaminonaphthalene/Pt diskChO, ChE
Entrapment
0.7 V versus Ag/AgCl15 sec97 MUp to 0.8 mM[41]

MWCN/screen printed carbon electrodeChO, ChE, HRP, K4Fe(CN)6
Physical adsorption
0.3 V versus Ag/AgCl180 sec100 mg/dL100–400 mg/dL2 months [27]

3-Aminopropyl-modified controlled-pore glass(APCEG)/rotating diskChO, ChE, HRP
Covalent cross-linking via Glutaraldehyde
−0.15 V versus Ag/AgCl with TBC as mediator11.9 nM1.2 M–1 mM25 days[42]

PANI/ITOChO, ChE
Covalent cross-linking via Glutaraldehyde
0.5 V versus Ag/AgCl40 sec50 mg/dL50–500 mg/dL6 weeks[43]

HRP incorporated carbon pasteChO, ChE
Covalent cross-linking on PVC beaker
−0.5 V versus Ag/AgCl20 sec2.5 mg/dL50–550 mg/dL100 days[38]

Nanoporous Au networks directly grown on a titanium substrateChO, ChE, HRP
Physical adsorption
Chitosan used as glue
Cyclic voltammetry0.5 mg/dL0.97–7.8 mM60 days[44]

ZnO–CuO composite matrix grown onto ITO coated corning glassChO, ChE
Physical adsorption
Cyclic voltammetry5 sec0.5 mM0.5–12 mM[45]

c-MWCNT/AuNPChO, ChE, HRP
Covalent cross-linking via c-MWCNT
0.27 V versus Ag/AgCl20 sec0.5 mg/dL0.5–300 mg/dL60 daysThis work