Ex Vivo Evaluation of Secretion-Clearing Device in Reducing Airway Resistance within Endotracheal Tubes

Table 2

Pressure drop characteristics of tested and control endotracheal tubes at 60 L/minute.

Type of endotracheal tubeMean pressure dropStandard deviationMin. pressure dropMax. pressure dropt-test value vs. control valuet-test value vs cleared tube value

All endotracheal tubes
All controls310.80.27n/a−8.63<0.001
Extubated, not cleared (pre)4021.813.3010.277.85.27<0.0014.80<0.001
Extubated, cleared (post)4011.80.299.516.8−8.63<0.001n/a
7.0 mm diameter
Control114.50.14n/a−3.690.035
Extubated, not cleared (pre)419.21.8816.420.44.940.0165.200.014
Extubated, cleared (post)415.90.7515.016.8−3.690.035n/a
7.5 mm diameter
Control111.80.01n/a−3.300.004
Extubated, not cleared (pre)1724.19.1212.046.04.33<0.0014.130.001
Extubated, cleared (post)1712.50.8611.414.3−3.300.004n/a
8.0 mm diameter
Control19.10.04n/a−10.76<0.001
Extubated, not cleared (pre)1923.017.4210.277.83.550.0023.230.004
Extubated, cleared (post)1910.30.489.511.3−10.76<0.001n/a