Pooled Analysis of the Efficacy and Safety of Video Capsule Endoscopy in Patients with Implantable Cardiac Devices

Table 2

Pooled results of the 16 studies included in the analysis [5ā€“20].

Type of DeviceTotal Number of cardiac devices (n)Cardiac devices brandVCE brandCardiac OutcomeQuality of transmitted images

PPM487Medtronic (77)
Guidant (54)
St Jude Medical (26)
Biotronik (12)
Vitatron (8)
ELA Sorin (3)
Boston scientific (1)
No specification (306)
PillCam (431)
EndoCapsule (33)
MiroCam (3)
No specification (20)
No adverse cardiac events
No cardiac device malfunction
1 case had some atrial and ventricular premature contractions without clinical symptoms
4 cases of VCE interference with telemetry leads leading to few images loss
1 case showed low capsule signal resulting in transient lack of localization of the capsule without images loss (PillCam with ELA Sorin)

ICD162Medtronic (15)
St Jude (13)
Guidant (10)
Boston scientific (7)
ELA Sorin (2)
Biotronik (1)
No specification (114)
PillCam (147)
EndoCapsule (8)
MiroCam (3)
No specification (4)
No adverse cardiac events
No cardiac device malfunction
1 case had some atrial and ventricular premature contractions without clinical symptoms
2 cases of VCE interference with telemetry leads leading to few images loss

LVAD86Thoratec HeartMate II (35)
HeartWare (3)
No specification (48)
PillCam (52)
No specification (34)
No adverse cardiac events
No cardiac device malfunction
5 cases of possible interference leading to images loss/artifact when VCE was near the device

Permanent Pacemaker (PPM); Implantable Cardiac Defibrillator (ICD); Left Ventricular Assist Device (LVAD); Video Capsule Endoscopy (VCE).