CT-Guided Percutaneous Microwave Ablation of Sclerosing Hepatic Carcinoma

Table 1

The average scope of the MWA probe in a freshly excised porcine liver using varying power and time at 37°C. The diameter of the MWA probe is 2 mm.

Power/timeTransverse diameter (mm)Vertical diameter (mm)Carbonized width (mm)Carbonized length (mm)

30 W/3 min23.8 ± 1.434.8 ± 1.75.5 ± 0.819.0 ± 1.8
30 W/5 min25.1 ± 1.239.0 ± 1.86.9 ± 0.620.8 ± 2.4
30 W/8 min26.0 ± 0.946.3 ± 2.17.1 ± 0.724.5 ± 3.3
40 W/3 min25.5 ± 1.536.8 ± 1.36.3 ± 0.819.6 ± 1.5
40 W/5 min26.2 ± 1.245.0 ± 2.67.5 ± 1.322.8 ± 2.8
40 W/8 min37.9 ± 2.153.8 ± 2.78.1 ± 0.926.5 ± 3.6
50 W/3 min33.8 ± 2.439.1 ± 1.56.5 ± 0.621.0 ± 2.3
50 W/5 min35.1 ± 1.949.2 ± 2.97.7 ± 0.727.8 ± 3.1
50 W/8 min46.0 ± 3.559.8 ± 3.89.2 ± 1.231.5 ± 3.5