CT-Guided Percutaneous Microwave Ablation of Sclerosing Hepatic Carcinoma
Table 1
The average scope of the MWA probe in a freshly excised porcine liver using varying power and time at 37°C. The diameter of the MWA probe is 2 mm.
| Power/time | Transverse diameter (mm) | Vertical diameter (mm) | Carbonized width (mm) | Carbonized length (mm) |
| 30 W/3 min | 23.8 ± 1.4 | 34.8 ± 1.7 | 5.5 ± 0.8 | 19.0 ± 1.8 | 30 W/5 min | 25.1 ± 1.2 | 39.0 ± 1.8 | 6.9 ± 0.6 | 20.8 ± 2.4 | 30 W/8 min | 26.0 ± 0.9 | 46.3 ± 2.1 | 7.1 ± 0.7 | 24.5 ± 3.3 | 40 W/3 min | 25.5 ± 1.5 | 36.8 ± 1.3 | 6.3 ± 0.8 | 19.6 ± 1.5 | 40 W/5 min | 26.2 ± 1.2 | 45.0 ± 2.6 | 7.5 ± 1.3 | 22.8 ± 2.8 | 40 W/8 min | 37.9 ± 2.1 | 53.8 ± 2.7 | 8.1 ± 0.9 | 26.5 ± 3.6 | 50 W/3 min | 33.8 ± 2.4 | 39.1 ± 1.5 | 6.5 ± 0.6 | 21.0 ± 2.3 | 50 W/5 min | 35.1 ± 1.9 | 49.2 ± 2.9 | 7.7 ± 0.7 | 27.8 ± 3.1 | 50 W/8 min | 46.0 ± 3.5 | 59.8 ± 3.8 | 9.2 ± 1.2 | 31.5 ± 3.5 |
|
|