[Retracted] Clinical Study on the Differential Diagnosis of High Myopia Astigmatism and Subclinical Keratoconus in Adolescents by Pentacam Anterior Segment Analyzer

Table 3

Analysis of the differential diagnosis value of various parameters for high myopia astigmatism and subclinical keratoconus.

ParameterAUC95% CI valueBest cutoff valueSensitivity (%)Specificity (%)

Thickness of the thinnest cornea0.5630.451∼0.6740.275512.3543.5632.35
Anterior vertex height of the thinnest point of the cornea0.8080.722∼0.8930.0015.5473.5874.56
Posterior vertex height of the thinnest point of the cornea0.8330.755∼0.9120.00110.5775.1670.37
IVA0.8680.796∼0.9400.0010.21679.5681.35
IHD0.8470.767∼0.9260.0010.01578.4475.23
Average corneal pachymetric progression index0.6840.578∼0.7900.0011.2060.3557.46
Joint diagnostic factor Y0.9260.863∼0.9890.0010.49586.9489.63