Research Article
FEM Analysis of Mandibular Prosthetic Overdenture Supported by Dental Implants: Evaluation of Different Retention Methods
Table 6
Maximum stresses for component [MPa] (Model C).
| COMP. | Equiv. elastic strain (max) | 0,027505 mm/mm | Midollar bone | Equivalent stress (max) | 206,47 MPa | Abutment implant |
| C. 45° | Equiv. elastic strain (max) | 0,040865 mm/mm | Midollar bone | Equivalent stress (max) | 635,08 MPa | Abutment implant |
| N. 90° | Equiv. elastic strain (max) | 0,047132 mm/mm | Midollar bone | Equivalent stress (max) | 856,52 MPa | Abutment implant |
| T. 45° | Equiv. elastic strain (max) | 0,041481 mm/mm | Midollar bone | Equivalent stress (max) | 732,56 MPa | Abutment implant |
| TRAZ. | Equiv. elastic strain (max) | 0,027519 mm/mm | Midollar bone | Equivalent stress (max) | 195,09 MPa | Cortical bone |
|
|