Research Article
Evaluation of Round Window Stimulation Performance in Otosclerosis Using Finite Element Modeling
Table 1
Dimensions of the human ear finite element model.
| Structure | Our model | Published data |
| Eardrum | | | Diameter along manubrium | 8.99 mm | 8.0–10.0 mm [47] | Diameter perpendicular to manubrium | 7.75 mm | 7.5–9.0 mm [47] | Thickness | 0.1 mm | 0.1 mm [47] | Malleus | | | Total length | 7.67 mm | 7.6–9.1 mm [47] | Incus | | | Length along long process | 7.10 mm | 7.0 mm [47] | Length along short process | 4.89 mm | 5.0 mm [47] | Stapes | | | Height | 3.75 mm | 2.5–4.0 mm [47] | Length of footplate | 2.95 mm | 2.64–3.36 mm [47] | Width of footplate | 1.51 mm | 0.7–1.66 mm [47] | Cochlea | | | Length of cochlea | 37 mm | 29.07–37.45 mm [48] | Length of BM | 34 mm | 35 mm [49] | Width of BM at base | 0.15 mm | 0.1 mm [49] | Width of BM at apex | 0.5 mm | 0.5 mm [49] | Thickness of BM at base | 7.5 μm | 7.5 μm [50] | Thickness of BM at apex | 2.5 μm | 2.5 μm [50] | Volume of scala vestibuli | 92.315 mm3 | 90.8463 mm3 [51] | Volume of scala tympani | 93.270 mm3 | 91.2044 mm3 [51] | Area of RWM | 2.1 mm2 | 2.08 mm2 [21] |
|
|