Research Article

Phantom Validation of Tc-99m Absolute Quantification in a SPECT/CT Commercial Device

Table 1

Background activity concentration (), total activity deviation (), calibration factor (Bg. cal), and coefficient of variation (COV) for 4 SPECT acquisitions of a cylindrical homogeneous phantom.

Homogeneous
scan number
(kBq/mL) (%)Bg. calCOV%

Scan 1223.91.0513
Scan 2204.11.0412
Scan 3183.51.0412
Scan 428.61.0938