Research Article

Phantom Validation of Tc-99m Absolute Quantification in a SPECT/CT Commercial Device

Table 3

SPECT acquisitions of the liver phantom: activity concentration in the background () and spherical inserts (), total activity deviation (), calibration factor (Bg. cal) and coefficient of variation (COV), and hot contrast () for the three inserts.

Liver
scan number
kBq/mL kBq/mL (%)Bg. calCOV% 40 mm 30 mm 20 mm

Scan 177.79404.571.91.088.47685130
Scan 262.13323.090.11.078.99675129
Scan 342.05218.70−0.11.0612.12675433
Scan 44.9925.955.81.0743.25715133