A Fetal Electrocardiogram Signal Extraction Algorithm Based on Fast One-Unit Independent Component Analysis with Reference

1College of Information and Communication Engineering, Harbin Engineering University, Heilongjiang 150001, China
2College of Electrical and Information Engineering, Beihua University, Jilin 132012, China
3Collaborative Research Center, Meisei University, Tokyo 1918506, Japan

Received 10 July 2016; Accepted 8 August 2016

Academic Editor: Po-Hsiang Tsui

Copyright © 2016 Yanfei Jia and Xiaodong Yang. This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Abstract

Fetal electrocardiogram (FECG) extraction is very important procedure for fetal health assessment. In this article, we propose a fast one-unit independent component analysis with reference (ICA-R) that is suitable to extract the FECG. Most previous ICA-R algorithms only focused on how to optimize the cost function of the ICA-R and payed little attention to the improvement of cost function. They did not fully take advantage of the prior information about the desired signal to improve the ICA-R. In this paper, we first use the kurtosis information of the desired FECG signal to simplify the non-Gaussian measurement function and then construct a new cost function by directly using a nonquadratic function of the extracted signal to measure its non-Gaussianity. The new cost function does not involve the computation of the difference between the function of the Gaussian random vector and that of the extracted signal, which is time consuming. Centering and whitening are also used to preprocess the observed signal to further reduce the computation complexity. While the proposed method has the same error performance as other improved one-unit ICA-R methods, it actually has lower computation complexity than those other methods. Simulations are performed separately on artificial and real-world electrocardiogram signals.