Research Article

Defining the Optimal Time of Adaptive Replanning in Prostate Cancer Patients with Weight Change during Volumetric Arc Radiotherapy: A Dosimetric and Mathematical Analysis Using the Gamma Index

Table 3

The changes in gamma index with changes in body contour from −2.0 cm to 2.0 cm.

Body contour changes (cm)
−2.0−1.5−1.0−0.500.51.01.52.0

6-MVGPR90.1091.8094.3098.00100.0098.7095.5093.8092.50
MG10.008.766.273.210.001.852.883.694.29
10-MVGPR91.6093.2095.7099.40100.0099.8096.9095.1093.30
MG6.894.813.191.750.001.392.192.853.30
15-MVGPR92.8094.6096.5099.70100.0099.8098.0095.8094.10
MG5.904.553.011.580.001.061.682.082.38

GPR, gamma pass rate; MG, maximum gamma values. Italic font: unacceptable.