Research Article
A Hybrid Wavelet-Based Method for the Peak Detection of Photoplethysmography Signals
Table 3
The total peaks identification results.
| Subject | Sitting | Raising hand | Gently walking | RB | Err_R (%) | Err_C (%) | RB | Err_R (%) | Err_C (%) | RB | Err_R (%) | Err_C (%) |
| (1) | 68 | 0.00 | 0.00 | 69 | 0.00 | 0.00 | 70 | 4.29 | 1.43 | (2) | 62 | 0.00 | 0.00 | 70 | 1.43 | 0.00 | 90 | 0.00 | 0.00 | (3) | 74 | 0.00 | 0.00 | 79 | 5.06 | 1.27 | 83 | 10.84 | 4.82 | (4) | 78 | 0.00 | 0.00 | 80 | 1.25 | 0.00 | 83 | 10.84 | 2.41 | (5) | 74 | 0.00 | 0.00 | 77 | 0.00 | 0.00 | 94 | 1.06 | 0.00 | (6) | 70 | 0.00 | 0.00 | 75 | 1.33 | 0.00 | 79 | 0.00 | 0.00 | (7) | 88 | 0.00 | 0.00 | 85 | 4.71 | 1.18 | 89 | 3.37 | 1.12 | (8) | 103 | 0.00 | 0.00 | 104 | 1.92 | 0.96 | 115 | 0.00 | 0.00 | (9) | 78 | 0.00 | 0.00 | 92 | 3.26 | 1.09 | 78 | 3.85 | 1.28 | (10) | 66 | 0.00 | 0.00 | 70 | 1.43 | 0.00 | 77 | 10.39 | 3.90 |
| Total | 761 | 0.00 | 0.00 | 801 | 2.12 | 0.50 | 858 | 4.20 | 1.40 |
|
|
Note. Err_R represents the detection error on the raw signals; Err_C means the detection error on the corrected signals.
|