Comparative Screening Analytic Methods for Elderly of Blood Methylmercury Concentration between Two Analytical Institutions
Table 1
The analytical conditions of instruments and recovery test result by instruments.
| | Total mercury | Methylmercury |
| Recovery test results |
| D Lab R(%RSD) | 101 % (0.4%) | 106 % (5.0%) |
| N Lab R(RSD) | 99 % (2.8%) | 101 % (1.4%) |
| Methylmercury analytic conditions of instruments | | |
| D Lab | Temperatures of the instruments were 65°C for digestion, 400~500°C for desorption, respectively. |
| N Lab GC conditions | Temperatures of the injection port, column oven and detector were 180°C, 160°C and 200°C, respectively. |
|
|
%R: recovery rate and %RSD: recovery standard deviation. Both Lab’ QC material was CRM (SRM 955c level 3). |