Research Article

Novel Signal Noise Reduction Method through Cluster Analysis, Applied to Photoplethysmography

Table 1

RMSE comparisons between noncluster and cluster algorithms for different noise situations. Note that where there is significant noise the cluster method returns a time-out error.

Panel in Figure 6Simulated noiseRMSE between clean median pulse and median of all noisy pulses (153 pulses)RMSE between clean median pulse and median of cluster (20 pulses)

Movement artefact (10% noise)0.030.14
Movement artefact (20% noise)0.120.14
Movement artefact (30% noise)0.230.11
Movement artefact (40% noise)0.270.07
(a)Movement artefact (50% noise)0.430.12
Movement artefact (60% noise)0.64
(b)Electrical noise0.120.09
(c)Movement (30%) and electrical noise0.66