Research Article
Numerical and Experimental Investigation of Novel Blended Bifurcated Stent Grafts with Taper to Improve Hemodynamic Performance
Table 2
Displacement forces (N) acting on two types of blended bifurcated stent graft (BBSG) under various degrees of the taper.
| | Conventional BBSG | Deflected BBSG |
| Taper (mm) | 17–10 | 17–9 | 17–8 | 17–10 | 17–9 | 17–8 | 60 (mmHg) | 0.78 ± 0.01 | 0.82 ± 0.01 | 0.87 ± 0.01 | 0.71 ± 0.01 | 0.74 ± 0.01 | 0.78 ± 0.01 | 80 (mmHg) | 0.93 ± 0.01 | 0.96 ± 0.01 | 1.01 ± 0.01 | 0.89 ± 0.01 | 0.92 ± 0.01 | 0.96 ± 0.01 | 100 (mmHg) | 1.12 ± 0.03 | 1.17 ± 0.01 | 1.24 ± 0.01 | 1.03 ± 0.01 | 1.07 ± 0.01 | 1.13 ± 0.01 |
|
|