Research Article

Finite Element- and Design of Experiment-Derived Optimization of Screw Configurations and a Locking Plate for Internal Fixation System

Table 6

Intuitionistic analysis for the orthogonal experiment of screw configurations.

Number of experimentsFactorsResults of experiment F
ABCDPNKL

1A1B1C1D1P1N1K1L1
2A1B2C2D2P2N2K2L2
3A1B3C3D3P3N3K3L3
4A2B1C2D3P4N4K4L4
5A2B2C3D1P5N5K5L5
6A2B3C1D2P6N6K6L6
7A3B1C3D2P7N7K7L7
8A3B2C1D3P8N8K8L8
9A3B3C2D1P9N9K9L9

I285.33262.33270.33250.33The maximum von Mises stress of the plate (P)
II282.33251.67231.00258.33
III201.67255.33268.00260.67
Range83.6710.6739.3310.33
Optimal schemeA3B2C2D1

I434.33389.00393.33375.67The maximum von Mises stress of the screw (N)
II392.33363.00387.67410.33
III372.67447.33418.33413.33
Range61.6784.3330.6737.67
Optimal schemeA3B2C2D1

I69.7776.2370.2374.40The maximum von Mises stress of the femur (K)
II80.2770.8775.0068.40
III71.0073.9375.8078.23
Range10.506.365.579.83
Optimal schemeA1B2C1D2

I7.718.277.137.54The maximum strain of the callus (L)
II7.727.388.147.94
III7.997.868.238.03
Range0.280.891.100.49
Optimal schemeA1B2C1D1