Research Article | Open Access

Volume 2021 |Article ID 5288844 | https://doi.org/10.1155/2021/5288844

Valeria Cantoni, Roberta Green, Carlo Ricciardi, Roberta Assante, Leandro Donisi, Emilia Zampella, Giuseppe Cesarelli, Carmela Nappi, Vincenzo Sannino, Valeria Gaudieri, Teresa Mannarino, Andrea Genova, Giovanni De Simini, Alessia Giordano, Adriana D’Antonio, Wanda Acampa, Mario Petretta, Alberto Cuocolo, "Comparing the Prognostic Value of Stress Myocardial Perfusion Imaging by Conventional and Cadmium-Zinc Telluride Single-Photon Emission Computed Tomography through a Machine Learning Approach", Computational and Mathematical Methods in Medicine, vol. 2021, Article ID 5288844, 8 pages, 2021. https://doi.org/10.1155/2021/5288844

Comparing the Prognostic Value of Stress Myocardial Perfusion Imaging by Conventional and Cadmium-Zinc Telluride Single-Photon Emission Computed Tomography through a Machine Learning Approach

Academic Editor: Rafik Karaman
Received08 Jun 2021
Revised30 Sep 2021
Accepted05 Oct 2021
Published16 Oct 2021

Abstract

We compared the prognostic value of myocardial perfusion imaging (MPI) by conventional- (C-) single-photon emission computed tomography (SPECT) and cadmium-zinc-telluride- (CZT-) SPECT in a cohort of patients with suspected or known coronary artery disease (CAD) using machine learning (ML) algorithms. A total of 453 consecutive patients underwent stress MPI by both C-SPECT and CZT-SPECT. The outcome was a composite end point of all-cause death, cardiac death, nonfatal myocardial infarction, or coronary revascularization procedures whichever occurred first. ML analysis performed through the implementation of random forest (RF) and -nearest neighbors (KNN) algorithms proved that CZT-SPECT has greater accuracy than C-SPECT in detecting CAD. For both algorithms, the sensitivity of CZT-SPECT (96% for RF and 60% for KNN) was greater than that of C-SPECT (88% for RF and 53% for KNN). A preliminary univariate analysis was performed through Mann-Whitney tests separately on the features of each camera in order to understand which ones could distinguish patients who will experience an adverse event from those who will not. Then, a machine learning analysis was performed by using Matlab (v. 2019b). Tree, KNN, support vector machine (SVM), Naïve Bayes, and RF were implemented twice: first, the analysis was performed on the as-is dataset; then, since the dataset was imbalanced (patients experiencing an adverse event were lower than the others), the analysis was performed again after balancing the classes through the Synthetic Minority Oversampling Technique. According to KNN and SVM with and without balancing the classes, the accuracy ( value = 0.02 and value = 0.01) and recall ( value = 0.001 and value = 0.03) of the CZT-SPECT were greater than those obtained by C-SPECT in a statistically significant way. ML approach showed that although the prognostic value of stress MPI by C-SPECT and CZT-SPECT is comparable, CZT-SPECT seems to have higher accuracy and recall.

1. Introduction

Risk stratification by noninvasive cardiac imaging has become increasingly important to optimize management and outcome in patients with coronary artery disease (CAD) [1]. Previous research indicated that stress single-photon emission computed tomography (SPECT) myocardial perfusion imaging (MPI) has been the most widely used nuclear cardiac imaging technique for the noninvasive assessment of cardiac disease, including the prognosis and choice of the most appropriate treatment strategies for patients with CAD [2]. Conventional- (C-) SPECT systems utilize sodium iodide crystals and parallel-hole collimators. This approach presents some technical limits; for instance, we can mention extended imaging time, low spatial resolution, and large doses of radiopharmaceuticals [3]. Recently, these limitations have been overcome with the introduction of gamma cameras with semiconductor cadmium-zinc-telluride (CZT) allowed to directly convert radiation into electric signals, bringing an improvement in image accuracy and acquisition time [4, 5].

Previous studies showed that CZT-SPECT findings can be used for risk stratification of patients referred to MPI for suspected or known CAD. Lima et al. [6] demonstrated that CZT-SPECT and C-SPECT provide similar prognostic results, with lower prevalence of hard events in patients with normal scan [6]. Yokota et al. [7] showed that the prognostic value of normal stress-only CZT-SPECT is at least comparable and may be even better than that of normal C-SPECT [7].

These biomedical technologies can produce big amount of data and, nowadays, different techniques have been used to obtain as much information as possible from data and signals [812]. Introducing machine learning (ML) in the healthcare sector can help clinicians in diagnosis and therapy planning, as well as in management of resources [13, 14]. Several studies have been conducted to test CAD detection using ML algorithms and to predict patient outcome [1518]. An innovative approach is to use ML models to compare the performance of biomedical technologies, and an evaluation of the performance in terms of diagnostic power has already been reported [19, 20], demonstrating CZT-SPECT has a better ability to detect CAD. To the best of our knowledge, the prognostic value of CZT-SPECT and C-SPECT has not been investigated to date by using ML techniques.

Therefore, the purposes of the present investigation were as follows: (1)To evaluate the prognostic value of C-SPECT and CZT-SPECT using ML-based approaches in patients with suspected or known CAD(2)To compare the prognostic performance of these biomedical instrumentations through ML

This use of ML—in this particular case, aimed at compering two biomedical technologies—represents, to authors’ best knowledge, one of the first attempts in literature.

2. Materials and Methods

2.1. Patients

Between February 2016 and May 2017, a total of 453 consecutive patients with suspected or known CAD were submitted by referring physicians to stress MPI for assessment of myocardial ischemia. For overall population, clinical history and cardiac risk factors were collected. Patients with a previous history of myocardial infarction, revascularization procedures, or a diagnosed atherosclerotic coronary disease were considered to have known CAD. The review committee of our institution approved the study (Protocol Number 110/17), and all patients gave informed consent.

2.2. Study Protocol

All patients were submitted to stress technetium-99m sestamibi-gated SPECT MPI by physical exercise or dipyridamole stress test, according to the recommendations of the European Association of Nuclear Medicine and European Society of Cardiology [21]. The protocol followed in this paper was the same employed in our previous research [20]. All patients underwent MPI by both C-SPECT and CZT-SPECT systems according to a randomized scheme in 1 : 1 ratio that determined which camera was used for first acquisition. For C-SPECT, a dual-head rotating gamma camera (E.CAM, Siemens Medical Systems, Hoffman Estates, IL, USA) was used. The acquisition time was 20 min for both stress and rest images. For CZT-SPECT (D-SPECT, Spectrum Dynamics, Caesarea, Israel), recordings were obtained using 9 pixilated CZT crystal detector columns mounted vertically spanning a 90 geometry. Scan duration was lower than 10 minutes for stress and lower than 5 minutes for rest imaging.

An automated software program (e-soft, 2.5, QGS/QPS, Cedars- Sinai Medical Center, Los Angeles, CA) was utilized to compute left ventricular (LV) volumes and ejection fraction (EF) and the scores incorporating both the extent and severity of perfusion defects, employing a standard segmentation of the 17 myocardial regions. The extent and grade of the quantitative defect were determined based on sex-specific normal limits while adding the scores of the 17 segments (from 0 for normal to 4 for absent perfusion) of the stress images allowed us to compute the summed stress score (SSS). A poststress LVEF greater than 45% and a SSS lower than 3 were considered normal.

2.3. Follow-Up Data

A follow-up questionnaire was collected by calling all patients by examinators blinded to patient’s test results. The outcomes evaluated as endpoints were all-cause death, cardiac death, nonfatal myocardial infarction, or coronary revascularization procedures whichever occurred first. Cardiac death occurred subsequently to acute myocardial infarction, congestive heart failure, and cardiac interventional procedure related. Myocardial infarction was recorded when chest pain or equivalent symptom complex, positive cardiac biomarkers, or typical electrocardiographic changes were reported [22]. The length of follow-up was determined according to the date of the last medical visit.

2.4. Statistical Analysis

Statistical analyses were performed by using IBM SPSS statistics software (v. 26), both to test data distribution and to perform statistical tests. The process was carried out separately on both the parameters of the C-SPECT and the CZT-SPECT. First, the Kolmogorov-Smirnov test was performed to test data normality, in order to understand the type of test to be used (parametric or nonparametric): in particular, normality was tested for all parameters, for both groups, and for both camera types. Subsequently, a two-tailed -test was performed for parameters with a normal distribution, while Mann-Whitney test was performed for the remaining parameters, and both tests were conducted considering a significance level of 0.05. After the use of ML algorithms, a chi-square test was used in order to compare the performances of different the models, trained with C-SPECT and the CZT-SPECT data, and to understand if there were statistical differences among them. The results are shown and discussed in the “Results” and “Discussion” sections, respectively.

2.5. Machine Learning Algorithms

The ML analysis was performed by using the Classification Learning App, provided by Matlab (v. 2019b), which trains models to classify data using supervised ML. The 10-fold crossvalidation was used to train and test the models; the dataset was divided into 10 groups of data, 9 were used for training the model and one group for testing it; the procedure was repeated 10 times, and the evaluation metrics are computed by averaging all those obtained [23]. The tree-based approach has shown in literature great results not only in the cardiologic context in cases such as diagnosis [2426], prognosis [27, 28], and comparison of biomedical technologies [19, 20] but also in other medical specialties [2931]. The classification tree is a simple and effective model consisting of nodes, branches, and leaves: each node has a rule that the data is routed along several branches while the leaves represent the output of the system [32].

Random forests (RF) model is part of the ensemble algorithms and allows to train together a set number of decision trees using the technique of Bootstrap Aggregation; this model turns out to have better accuracy than the single weak learner and reduces the chance of overfitting [33]. -nearest neighbor (KNN) algorithm is a distance-based method. In fact, an example’s membership in a class is determined by proximity to other known class examples. The critical aspect is the choice of the value of that is the number of neighbors to consider for the decision [34]. Support vector machine (SVM) is a classification model that is based on finding the best surface that allows you to separate the two classes. In particular, the algorithm tries to maximize the margin between classes, the space that separates them, and in this way, bases learning on the most difficult examples, decreasing the influence of outliers [35]. Naïve Bayes (NB) was also employed in this study; it is a well-known algorithm based on the a priori probability theorem [36], thus being a completely different algorithm compared to the previous ones. These algorithms were used to predict an adverse event by using the features of the two cameras, and then, the evaluation metrics were compared through a statistical test for proportions in order to understand which one had the best capacity to detect the adverse event.

The present dataset is unbalanced; indeed, people with adverse events turned out to be much less than those with no events. In the literature, the problems that arise in training ML models using unbalanced data are well known [37, 38]. To deal with this problem, the Synthetic Minority Oversampling Technique (SMOTE) [39] was used; this oversampling technique generates new artificial data of the minority class, on the basis of those already present, allowing to rebalance the dataset. After that, the training phase of the models was repeated. This can be considered fair because it will be employed on both cameras allowing a fair comparison; moreover, the aim of the study is to compare C-SPECT and the CZT-SPECT rather than build the best prognostic model.

To evaluate the performance of the models, several metrics [40] were used: accuracy, sensitivity or recall, specificity, and precision. Furthermore, area under the curve (AUC) receiver-operating characteristic (ROC) was computed because it is a good method to assess model performance [41]. In addition, a feature selection process was performed to understand which parameters resulted more significant in reference to the target variable. We tested 14 features: perfusion parameters as SSS, summed rest score (SRS), summed difference score (SDS), and total perfusion defect (TPD) and functional parameters as systolic wall motion (SWM), systolic wall thickening (SWT), end-diastolic volume (EDV), end-systolic volume (ESV), and EF. In particular, two algorithms were used: Maximum Relevance–Minimum Redundancy (MRMR) that selects the variables with the most relevance to the destination one by calculating the mutual information of the parameters [42] and chi-square independence test [43].

3. Results

3.1. Patient Characteristics and Outcome

The clinical characteristics of patient population are shown in Table 1. The study group comprised 204 (45%) patients with suspected CAD and 249 (55%) with known CAD. The mean follow-up was . During follow-up, 41 events occurred. The events were cardiac death in 1 patient, nonfatal myocardial infarction in 5, coronary revascularization procedures in 20, and 15 all-cause of death.


Characteristic

Age (years)
Male gender, (%)331 (73)
, (%)110 (24)
Diabetes, (%)153 (34)
Dyslipidemia, (%)333 (74)
Smoking, (%)196 (43)
Hypertension, (%)386 (85)
Atypical angina, (%)162 (36)
Family history of CAD, (%)231 (51)
Previous myocardial infarction, (%)148 (33)
Previous revascularization procedures, (%)173 (38)

Data are presented as or (%) of subjects. CAD: coronary artery disease.
3.2. Statistical Analysis

The first step was to evaluate the possible normal distribution of the features between patients with events and patients with no events evaluated by both cameras, applying Kolmogorov-Smirnov test. The test revealed that, among the features of C-SPECT, only stress and rest EF ( value > 0.05) showed a normal distribution for both groups; similarly, no features of CZT-SPECT resulted to have a Gaussian distribution. Therefore, -test was used only for stress and rest EF by C-SPECT, while Mann-Whitney test was performed for all other parameters, and the results are reported in Table 2.


ParametersC-SPECTCZT-SPECT
Patients with no eventPatients with event valuePatients with no eventPatients with event value

SSS0.053<0.001
SRS0.1630.240
SDS0.8410.310
TPD0.043<0.001
Stress SWM0.0070.018
Stress SWT0.0020.020
Stress EDV0.0310.044
Stress ESV0.0060.008
Stress EF<0.0010.005
Rest SWM0.0480.070
Rest SWT0.1280.110
Rest EDV0.4930.790
Rest ESV0.2830.420
Rest EF0.0980.160

Statistically significant at: 0.05, 0.001, <0.001. Abbreviations. EDV: end-diastolic volume; EF: ejection fraction; ESV: end-systolic volume; SDS: summed difference score; SRS: summed rest score; SSS: summed stress score; SWM: wall motion; SWT: wall thickening; TPD: total perfusion defect.
3.3. Machine Learning Analysis

The ML analysis was conducted separately and by using a 10-fold crossvalidation for C-SPECT and CZT-SPECT, both before and after SMOTE application in order to compare camera’s performance with and without the augmentation of the dataset. The evaluation metrics regarding the models without SMOTE are reported in Table 3. Among the ML algorithms used for the analysis, RF reached the highest value of accuracy (90.3% and 90.1%, respectively, for C-SPECT and CZT-SPECT) and recall (98.5% and 99.0%, respectively, for C-SPECT and CZT-SPECT), but it presented the lowest value of specificity (7.3% and 0%, respectively, for C-SPECT and CZT-SPECT), showing a low capacity to detect adverse future events. Despite achieving these performances, statistically significant differences between the two cameras were not available, and this was also verified for Tree, SVM, and NB models. KNN model had an accuracy and recall lower than RF for both cameras (accuracy of 74.4% and 80.8%, recall of 78.6% and 87.4%, respectively, for C-SPECT and CZT-SPECT) but higher specificity (ranging from 14.6%, in CZT-SPECT, to 31.7% in C-SPECT). Nevertheless, the accuracy and the capacity to detect the absence of adverse event were statistically significant in favour of the CZT camera ( value = 0.021 for accuracy and value = 0.001 for recall). These results were influenced by the imbalanced nature of the datasets; indeed, although accuracy and recall were high, they were affected by the bias introduced by the presence of a majority class for subjects with a negative prognosis, as also validated by the low AUCROC values of the models (ranging from 0.53 to 0.60 for C-SPECT and from 0.50 to 0.61 for CZT-SPECT). To overcome this issue, the dataset was balanced by introducing artificial samples of the minority class (patients with future adverse events), generated with SMOTE. The evaluation metrics values are reported in Table 4. The overall performance of classifiers increased significantly with a balanced dataset, especially in terms of specificity and AUCROC. Considering the C-SPECT, RF reached the highest values of accuracy (93.4%), recall (90.3%), and AUCROC (0.99), while SVM and KNN reached higher values of specificity (95.0% and 99.8%, respectively). Regarding CZT camera models performances, SVM classifier reached the highest values of accuracy (94.5%), recall (92.2%), and specificity (96.8%). Moreover, SVM turned out to have statistically significant performances: the accuracy and the recall showed a statistical significance in favor of the CZT-camera ( value = 0.016 for accuracy and value = 0.028 for recall), while, despite showing in CZT-SPECT a higher capacity to detect adverse events, the specificity of SVM was not found to be statistically significant ( value = 0.279).


Accuracy (%)Error (%)Recall (%)Specificity (%)

TreeC-SPECT87.412.694.417.1
CZT-SPECT89.011.097.17.32
value0.4710.0570.177
KNNC-SPECT74.425.678.631.7
CZT-SPECT80.819.287.414.6
value0.0210.0010.067
SVMC-SPECT85.914.192.221.6
CZT-SPECT86.513.592.621.6
value0.7730.5971.000
NBC-SPECT83.416.689.126.8
CZT-SPECT84.115.990.124.4
value0.7870.6490.800
RFC-SPECT90.39.798.57.3
CZT-SPECT90.19.999.00.0
value0.5910.5250.078

Abbreviations: KNN: nearest neighbor; SVM: support vector machine; NB: Naïve Bayes; RF: random forests.

Accuracy (%)Error (%)Recall (%)Specificity (%)

TreeC-SPECT88.111.986.290.1
CZT-SPECT88.111.986.989.3
value1.0000.7600.731
KNNC-SPECT91.98.183.999.8
CZT-SPECT91.68.484.798.5
value0.8580.7740.058
SVMC-SPECT91,58.587,695.0
CZT-SPECT94.55.592.296.8
value0.0160.0280.279
NBC-SPECT59.340.786.732.0
CZT-SPECT59.041.087.630.3
value0.8800.6770.599
RFC-SPECT93.46.690.394.4
CZT-SPECT93.07.091.094.9
value0.6370.7200.757

Abbreviations. KNN: nearest neighbor; SVM: support vector machine; NB: Naïve Bayes; RF: random forests.

4. Discussion

To our knowledge, this is the first study using ML approach to compare the prognostic value of two technologies used in clinical routine practice (C-SPECT and CZT-SPECT) in patients with suspected or known CAD. Indeed, the ML analysis did not aim to create the best model to predict adverse events because, probably, it would not have been possible considering the highly unbalanced nature of the dataset. The aim was to test the feasibility of the cameras in predicting adverse future events in order to understand which could be the one with the better performance.

Although a similar evaluation has already been performed, ML techniques have never been used. Lima et al. [6] compared the prognostic value of MPI using an ultrafast protocol with low radiation in CZT-SPECT and a C-SPECT in different groups of patients. They concluded that the new protocol of MPI in CZT-SPECT showed similar prognostic results to those obtained in dedicated cardiac Na-I SPECT camera, with lower prevalence of hard events in patients with normal scan. Similarly, Yokota et al. [7] compared the prognosis of patients with normal stress-only at both CZT-SPECT and C-SPECT. They showed that the prognostic value of normal stress-only CZT-SPECT is at least comparable and may be even better than that of normal stress-only C-SPECT. In a recent study, Liu et al. [44] showed that ultralow dose thallium perfusion imaging using CZT-SPECT provides good prognostic results, with a more severe prognosis in patients with abnormal MPI.

However, ML has been recently employed for the comparison of biomedical technologies. In previous studies using ML techniques to compare the diagnostic performance of C-SPECT and CZT-SPECT, we highlighted how algorithms trained with CZT-SPECT data achieved better accuracy, recall, and specificity than C-SPECT [19, 20]. Concerning the ML models, it has been observed that they generally present a high accuracy and recall. In particular, accuracy ( value = 0.021) and recall ( value = 0.001) were statistically significant for CZT-SPECT through the KNN algorithm. This result would demonstrate that CZT-SPECT has better performance to detect the absence of adverse event. To enhance the results obtained on the unbalanced dataset, a process of rebalancing the dataset was applied using SMOTE and repeating all the ML analyses. As expected, the performance of all models improved significantly for both cameras after rebalancing. However, SVM showed marked differences in all metrics values: accuracy, recall, and specificity had higher values in CZT-SPECT than C-SPECT. In particular, accuracy and recall were statistically significant in favour of CZT-SPECT (accuracy value = 0.016, recall value = 0.028). Therefore, even considering the balanced data, CZT-SPECT proved to achieve a better accuracy and ability in predicting the absence of adverse event. It is likely that patients affected by an adverse event had a particular pattern of input variables which have allowed instance-based algorithms (KNN and SVM) to capture the outcome better than tree-based and probability-based algorithms. As regards the computational costs and the runtime of our models, there was no specific problem because all the models followed a simple workflow without applying heavy preprocessing algorithms (such as backward or forward feature selection methods). Indeed, all the models required less than a minute to be run.

The novel CZT technique provides patients with several advantages, as lower radiation dose and imaging time. Moreover, the higher energy and intrinsic spatial resolution of CZT detectors lead to lower artifacts and need for rest imaging, with a consequent reduction in radiopharmaceutical dosage which enables nuclear MPI to be more cost-effective [45].

4.1. Limitations and Future Developments

This study has some limitations that need to be considered. The dataset was strongly imbalanced, to the detriment of patients who present adverse events. It influenced the learning process of the models, introducing biases into evaluation metrics. SMOTE technique has been applied to balance the dataset and overcome these issues. However, the samples introduced were artificial, which represented another limitation. Nevertheless, the aim of the paper was not to evaluate the performance of the models in order to create a tool for clinical support, but to compare the performance of two technologies; therefore, the limitation introduced by the oversampling process is attenuated. Regarding future developments, it would be necessary to try to balance the dataset with original data rather than with artificial samples in order to increase the reliability of the evaluation metrics.

5. Conclusions

The novelty introduced in this study was the use of supervised learning techniques to compare the prognostic value of C-SPECT and CZT-SPECT. The results obtained showed that although the prognostic value of the two systems is comparable; CZT-SPECT seems to have higher accuracy and recall.

Data Availability

The dataset used to support the findings of this study have not been made available because of the privacy policy.

Ethical Approval

All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki Declaration and its later amendments or comparable ethical standards.

Conflicts of Interest

The authors declare that they have no conflict of interests.

References

  1. L. J. Shaw and A. E. Iskandrian, “Prognostic value of gated myocardial perfusion SPECT,” Journal of Nuclear Cardiology, vol. 11, no. 2, pp. 171–185, 2004. View at: Publisher Site | Google Scholar
  2. W. Acampa, M. Petretta, L. Evangelista et al., “Stress cardiac single-photon emission computed tomographic imaging late after coronary artery bypass surgery for risk stratification and estimation of time to cardiac events,” Journal of Thoracic and Cardiovascular Surgery, vol. 136, no. 1, pp. 46–51, 2008. View at: Publisher Site | Google Scholar
  3. H. O. Anger, “Scintillation camera with multichannel collimators,” Journal of Nuclear Cardiology, vol. 5, pp. 515–531, 1964. View at: Google Scholar
  4. T. Sharir, S. Ben-Haim, K. Merzon et al., “High-speed myocardial perfusion imaging: initial clinical comparison with conventional dual detector anger camera imaging,” JACC: Cardiovascular Imaging, vol. 1, no. 2, pp. 156–163, 2008. View at: Publisher Site | Google Scholar
  5. D. S. Berman, X. Kang, B. Tamarappoo et al., “Stress thallium-201/rest technetium-99m sequential dual isotope high-speed myocardial perfusion imaging,” JACC Journals of the American College of Cardiology, vol. 2, no. 3, pp. 273–282, 2009. View at: Publisher Site | Google Scholar
  6. R. Lima, T. Peclat, T. Soares, C. Ferreira, A. C. Souza, and G. Camargo, “Comparison of the prognostic value of myocardial perfusion imaging using a CZT-SPECT camera with a conventional anger camera,” Journal of Nuclear Cardiology, vol. 24, no. 1, pp. 245–251, 2017. View at: Publisher Site | Google Scholar
  7. S. Yokota, M. Mouden, J. P. Ottervanger et al., “Prognostic value of normal stress-only myocardial perfusion imaging: a comparison between conventional and CZT-based SPECT,” European Journal of Nuclear Medicine and Molecular Imaging, vol. 43, no. 2, pp. 296–301, 2016. View at: Publisher Site | Google Scholar
  8. P. Bifulco, M. Cesarelli, L. Loffredo, M. Sansone, and M. Bracale, “Eye movement baseline oscillation and variability of eye position during foveation in congenital nystagmus,” Documenta Ophthalmologica, vol. 107, no. 2, pp. 131–136, 2003. View at: Publisher Site | Google Scholar
  9. M. Romano, F. Clemente, G. D'Addio, A. M. Ponsiglione, G. Improta, and M. Cesarelli, “Symbolic dynamic and frequency analysis in foetal monitoring,” in Proceedings of the IEEE International Symposium on Medical Measurements and Applications (MeMeA), 2014. View at: Google Scholar
  10. M. Cesarelli, A. Fratini, P. Bifulco, A. La Gatta, M. Romano, and G. Pasquariello, “Analysis and modelling of muscles motion during whole body vibration,” Eurasip Journal on Advances in Signal Processing, vol. 2010, no. 1, 2009. View at: Publisher Site | Google Scholar
  11. V. Onesto, L. Cancedda, M. L. Coluccio et al., “Nano-topography enhances communication in neural cells networks,” Scientific Reports, vol. 7, no. 1, 2017. View at: Publisher Site | Google Scholar
  12. M. Recenti, C. Ricciardi, K. Edmunds, M. K. Gislason, and P. Gargiulo, “Machine learning predictive system based upon radiodensitometric distributions from mid-thigh CT images,” European Journal of Translational Myology, vol. 30, no. 1, pp. 121–124, 2020. View at: Publisher Site | Google Scholar
  13. H. C. Koh and G. Tan, “Data mining applications in healthcare,” Journal of Health Informatics & Management, vol. 19, pp. 64–72, 2005. View at: Google Scholar
  14. J. Waring, C. Lindvall, and R. Umeton, “Automated machine learning: review of the state-of-the-art and opportunities for healthcare,” Artificial Intelligence in Medicine, vol. 104, p. 101822, 2020. View at: Publisher Site | Google Scholar
  15. A. Lin, M. Kolossváry, M. Motwani et al., “Artificial intelligence in cardiovascular imaging for risk stratification in coronary artery disease,” Radiology: Cardiothoracic Imaging, vol. 3, no. 1, 2021. View at: Publisher Site | Google Scholar
  16. A. Akella and S. Akella, “Machine learning algorithms for predicting coronary artery disease: efforts toward an open source solution,” Future Science OA, vol. 7, no. 6, 2021. View at: Publisher Site | Google Scholar
  17. M. Abdar, W. Ksiazek, R. Archarya, R. Tan, V. Makarenkov, and P. Plawiak, “A new machine learning technique for an accurate diagnosis of coronary artery disease,” Computer Methods and Programs in Biomedicine, vol. 179, p. 104992, 2019. View at: Publisher Site | Google Scholar
  18. J. Betancur, Y. Otaki, M. Motwani et al., “Prognostic value of combined clinical and myocardial perfusion imaging data using machine learning,” JACC Journals of the American College of Cardiology, vol. 11, no. 7, pp. 1000–1009, 2018. View at: Publisher Site | Google Scholar
  19. T. Mannarino, R. Assante, C. Ricciardi et al., “Head-to-head comparison of diagnostic accuracy of stress-only myocardial perfusion imaging with conventional and cadmium-zinc telluride single-photon emission computed tomography in women with suspected coronary artery disease,” Journal of Nuclear Cardiology, vol. 20, 2021. View at: Publisher Site | Google Scholar
  20. V. Cantoni, R. Green, C. Ricciardi et al., “A machine learning-based approach to directly compare the diagnostic accuracy of myocardial perfusion imaging by conventional and cadmium-zinc telluride SPECT,” Journal of Nuclear Cardiology, vol. 18, 2020. View at: Publisher Site | Google Scholar
  21. Task Force Members, G. Montalescot, U. Sechtem et al., “2013 ESC guidelines on the management of stable coronary artery disease,” European Journal of Nuclear Medicine and Molecular Imaging, vol. 34, no. 38, pp. 2949–3003, 2013. View at: Publisher Site | Google Scholar
  22. K. Thygesen, J. S. Alpert, A. S. Jaffe, M. L. Simoons, B. R. Chaitman, and H. D. White, “Third universal definition of myocardial infarction,” Circulation, vol. 126, no. 16, pp. 2020–2035, 2012. View at: Publisher Site | Google Scholar
  23. R. Kohavi, “A study of cross-validation and bootstrap for accuracy estimation and model selection,” Ijcai, IJCAI, vol. 14, no. 2, pp. 1137–1145, 1995. View at: Google Scholar
  24. C. Ricciardi, V. Cantoni, G. Improta et al., “Application of data mining in a cohort of Italian subjects undergoing myocardial perfusion imaging at an academic medical center,” Computer Methods and Programs in Biomedicine, vol. 189, p. 105343, 2020. View at: Publisher Site | Google Scholar
  25. C. Wang, Y. Zhao, B. Jin et al., “Development and validation of a predictive model for coronary artery disease using machine learning,” Frontiers in Cardiovascular Medicine, vol. 8, 2021. View at: Publisher Site | Google Scholar
  26. C. Ricciardi, A. S. Valente, K. Edmund et al., “Linear discriminant analysis and principal component analysis to predict coronary artery disease,” Health Informatics Journal, vol. 26, no. 3, pp. 2181–2192, 2020. View at: Publisher Site | Google Scholar
  27. C. Ricciardi, V. Cantoni, R. Green, G. Improta, and M. Cesarelli, “Is it possible to predict cardiac death?” in Mediterranean Conference on Medical and Biological Engineering and Computing, pp. 847–854, Springer, 2019. View at: Google Scholar
  28. C. Ricciardi, K. J. Edmunds, M. Recenti et al., “Assessing cardiovascular risks from a mid-thigh CT image: a tree-based machine learning approach using radiodensitometric distributions,” Scientific Reports, vol. 10, no. 1, 2020. View at: Publisher Site | Google Scholar
  29. A. Stanzione, C. Ricciardi, R. Cuocolo et al., “MRI radiomics for the prediction of Fuhrman grade in clear cell renal cell carcinoma: a machine learning exploratory study,” Journal of Digital Imaging, vol. 33, no. 4, pp. 879–887, 2020. View at: Publisher Site | Google Scholar
  30. C. Ricciardi, R. Cuocolo, G. Cesarelli et al., “Distinguishing functional from non-functional pituitary macroadenomas with a machine learning analysis,” Paper presented at the IFMBE Proceedings, vol. 76, pp. 1822–1829, 2020. View at: Google Scholar
  31. D. Scrutinio, C. Ricciardi, L. Donisi et al., “Machine learning to predict mortality after rehabilitation among patients with severe stroke,” Scientific Reports, vol. 10, no. 1, pp. 1–10, 2020. View at: Publisher Site | Google Scholar
  32. L. Breiman, J. Friedman, R. Olshen, and C. J. Stone, Classification and Regression Trees, IJCAI, 1983.
  33. L. Breiman, “Random forests,” Machine Learning, vol. 45, no. 1, pp. 5–32, 2001. View at: Publisher Site | Google Scholar
  34. S. Zhang, X. Li, M. Zong, X. Zhu, and R. Wang, “Efficient kNN classification with different numbers of nearest neighbors,” IEEE Transactions on Neural Networks and Learning Systems, vol. 29, no. 5, pp. 1774–1785, 2018. View at: Publisher Site | Google Scholar
  35. C. Cortes and V. Vapnik, “Support-vector networks,” Machine Learning, vol. 20, no. 3, pp. 273–297, 1995. View at: Publisher Site | Google Scholar
  36. I. Rish, “An empirical study of the naive Bayes classifier,” in IJCAI 2001 workshop on empirical methods in artificial intelligence, vol. 3, no. 22, pp. 41–46, 2001. View at: Google Scholar
  37. N. V. Chawla, N. Japkowicz, and A. Kotcz, “Editorial,” Special Interest Group on Knowledge Discovery in Data, vol. 6, no. 1, pp. 1–6, 2004. View at: Publisher Site | Google Scholar
  38. S. Kotsiantis, D. Kanellopoulos, and P. Pintelas, “Handling imbalanced datasets: a review,” GESTS international Transactions on Computer Science and Engineering, vol. 30, 2006. View at: Google Scholar
  39. N. V. Chawla, K. W. Bowyer, L. O. Hall, and W. P. Kegelmeyer, “SMOTE: synthetic minority over-sampling technique,” Journal of Artificial Intelligence Research, vol. 16, pp. 321–357, 2002. View at: Publisher Site | Google Scholar
  40. M. Hossin and M. N. Sulaiman, “A review on evaluation metrics for data classification evaluations,” International Journal of Data Mining & Knowledge Management Process, vol. 5, no. 2, pp. 01–11, 2015. View at: Publisher Site | Google Scholar
  41. A. P. Brandley, “The use of the area under the ROC curve in the evaluation of machine learning algorithms,” Pattern Recognition, vol. 30, no. 7, pp. 1145–1159, 1997. View at: Publisher Site | Google Scholar
  42. C. Ding and H. Peng, “Minimum redundancy feature selection from microarray gene expression data,” Journal of Bioinformatics and Computational Biology, vol. 3, no. 2, pp. 185–205, 2005. View at: Publisher Site | Google Scholar
  43. A. Satorra and P. M. Bentler, “A scaled difference chi-square test statistic for moment structure analysis,” Psychometrika, vol. 66, no. 4, pp. 507–514, 2001. View at: Publisher Site | Google Scholar
  44. L. Liu, F. A. Abdu, G. Yin et al., “Prognostic value of myocardial perfusion imaging with D-SPECT camera in patients with ischemia and no obstructive coronary artery disease (INOCA),” Journal of Nuclear Cardiology, 2020. View at: Publisher Site | Google Scholar
  45. W. Acampa, R. R. Buechel, and A. Gimelli, “Low dose in nuclear cardiology: state of the art in the era of new cadmium-zinc-telluride cameras,” European Journal of Nuclear Medicine and Molecular Imaging, vol. 17, no. 6, pp. 591–595, 2016. View at: Publisher Site | Google Scholar

Copyright © 2021 Valeria Cantoni et al. This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.