Case Report

Retrograde Intramedullary Nailing with a Blocking Pin Technique for Reduction of Periprosthetic Supracondylar Femoral Fracture after Total Knee Arthroplasty: Technical Note with a Compatibility Chart of the Nail to Femoral Component

Table 1

Intercondylar notch designs of the inserted TKAs and their compatibility with supracondylar nails. All these dimensions were officially provided by the manufactures.

Manufacture & model name of the componentSize of the componentIntercondylar width of the component (mm)Distance from the anterior femoral osteotomy phase to the open box anterior edge of the component (mm)

Senko Medical (Tokyo, Japan)
 Quest CR1–515–18.618.9–24.7
Smith & Nephew (Memphis, TN)
 Genesis II CR1–816.5–1916–19.9
 Genesis II PS1–616.511–23
 Legion CR1–816.5–1916–19.9
 Legion PS1–616.59.5–17
 Profix CRAll20N/A
 Profix PSAll15N/A
Zimmer (Warsaw, IN)
 Nexgen CR-FlexA–G11.9–13.318.9–27
 Nexgen LPSA–G13.7–21.215.3–20
 Nexgen LPS-FlexA–G13.7–21.215.3–20
 MG I CRS–large plus11–14N/A
 MG II CRAll12N/A
 IB I PSAll16N/A
 IB II PS54–6415–18N/A
Stryker (Kalamazoo, MI)
 Scorpio CR, PS3–1316.5–20.58–17.1
 NRG CR3–1318–22.215.8–26.8
 NRG PS3–1318–22.25.9–17.1
 Kinemax Plus CR, PSX small–large17–219.9–21.6
 Omnifit 3000 CR, PS3–1119.5N/A
 Delta Fit 7000 CR, PS3–1319.517.9–20.8
 Triathlon CR1–716.110–16.5
 Triathlon PS1–720.810–16.5
Depuy (Warsaw, IN)
 Sigma CR1.5–517.812–16.7
 LCS RP CRSmall–large plus14.4–21.912.7–20.9
Biomet (Warsaw, IN)
 ACG CR55–7518.1–23.9N/A
 Maxim CR55–7513.3–15.3N/A
 Maxim PS55–7515.2N/A
 Vanguard CR55–7513.3–15.313.9–20.1
 Vanguard PS55–7516.310.8–17.7
 Vanguard RP CR, PS55–7016.33–3.8
Kyocera Medical (Osaka, Japan)
 Bisurface PSX small–X large16.54.5–9.5
 LFA CR1–418–22.519.5–24.5
 LFA PS1–4189.4–14.4

CR, cruciate retaining; PS, posterior stabilized; N/A, not available.