Intermittent Abdominal Pressure Ventilation: An Alternative for Respiratory Support

Table 2

IAPV parameters: we suggest starting with Pbelt of 0–70 Hpa (at the beginning: 30–40 Hpa); select desired Ti (during Ti set, PBAir will be deflated, while the patient will be able to inhale); backup rate as desired; rise time usually 1.0 s; expiratory time (abdominal compression) will be linked to the backup rate and inspiratory time set. For example, set inspiratory time: 1.5 sec, Fr: 15 bpm, and derivative expiratory time: 2.5 sec.

Intermittent abdominal pressure ventilator (LunaBelt)
ModeTimedSpontaneous/timed

Pressure belt0–70 hPa0–70 hPa
Time inspiratory0.3–5.0 secNa
Time inspiratory minimumNa0.3–3.0 sec
Time inspiratory maximum[(60/Freq) − 0.6 sec][(60/Freq − 0.6 sec)]
Time expiratory minimumNa0–1.5 sec
Backup frequency1–60 bpm1–60 bpm
Frequency maximum[60/(Tinsp + 0.6 sec)][60/(Tinsp + 0.6 sec)
Rise time0.1–1.0 sec0.1–1.0 sec
Trigger inspiratory (nasal cannula)NaAuto
Trigger expiratory (nasal cannula)NaAuto