Research Article

Comparison of Actual Performance in the Flow and Fraction of Inspired O2 among Different High-Flow Nasal Cannula Devices: A Bench Study

Table 2

The structure and mechanism differences among devices.

DevicesAir supplyaOxygen supplybAir-oxygen mixercFiO2 monitord

AIRVO 2Integrated turbinesLow-pressure oxygen supply from separated oxygen flowmetersManually adjusts oxygen flow to titrate the aimed FiO2 based on monitored or calculated FiO2 in real timeUltrasonic oxygen analyzer
TNI softFlow 50Not mentioned in public information
HUMID-BH
OH-70CWall supplyProportional valvesFiO2 monitor provided
Bellavista 1000Wall supply

Note: aAir supply may be responsible for flow accuracy. bOxygen supply would affect the FiO2 accuracy. Flowmeters cannot provide enough oxygen flow when higher FiO2 at high-flow levels is needed. cChanges in settings would alter the input air and oxygen flow or pressure, which will challenge the air-oxygen mixer and cause an inaccurate output. dThe FiO2 monitor influences the FiO2 accuracy.