Comparison of Actual Performance in the Flow and Fraction of Inspired O2 among Different High-Flow Nasal Cannula Devices: A Bench Study

Table 4

Characteristics of the flow change rate in five devices.

DevicesFlow change rate (%)
≤35 L/min40–60 L/min>60 L/min

TNI softFlow 500 (−0.3, 0)0 (0, 0.2)N/A
Bellavista 10003.9 (3.6, 4.1)5.4 (5, 7.9)28.2 (22.6, 31.9)
HUMID-BH6.9 (4.2, 10.2)4.3 (3.2, 7.5)30.6 (23.8, 38)
OH-70C5.8 (4.1, 8.6)14.4 (6.3, 26.9)39.3 (37.7, 42.3)
AIRVO 28.3 (7.1, 9.9)31.4 (23.3, 38.3)N/A

The data are presented as the median (IQR). The flow change rate refers to the change in flow after one side of the cannulas is clipped.