Radiation Dose Optimization in Interventional Cardiology: A Teaching Hospital Experience
Table 6
Quantitative image quality results using the CIRS model 901 30 cm phantom for standard NEMA measurements.
| | Phase 1 | Phase 2 | Phase 3 | Phase 4 | Phase 5 | Fluoroscopy | Acquisitions | Fluoroscopy | Acquisitions | Fluoroscopy | Acquisitions | Fluoroscopy | Acquisitions | Fluoroscopy | Acquisitions |
| X-ray parameters | kVp | 118 | 90 | 113 | 90 | 110 | 90 | 110 | 102 | 110 | 102 | Tube current (mA) | 11.4 | 746 | 11 | 746 | 7.1 | 746 | 7.1 | 341 | 7.1 | 340 | Dose setting | Normal | Normal | Low | Normal | Low | Normal | Low | Normal | Low | Low | Frame rate | 7.5 | 15 | 7.5 | 15 | 5 | 15 | 5 | 10 | 5 | 10 | Added filteration | Al 2.0 + Cu 0.1 | Al 1.0 + Au 0.01 | Al 1.5 + Cu 0.6 | Al 1.0 + Au 0.01 | Al 1.5 + Cu 0.6 | Al 1.0 + Au 0.01 | Al 1.5 + Cu 0.6 | Al 1.0 + Au 0.01 | Al 1.5 + Cu 0.6 | Al 2.0 + Cu 0.1 |
| Static image quality | Line pairs | 1.6 | 1.8 | 1.4 | 1.8 | 1.2 | 1.8 | 1.2 | 1.6 | 1.2 | 1.6 | Iodine group 1 | 4 | 6 | 4 | 6 | 2 | 6 | 2 | 6 | 2 | 6 | Iodine group 2 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | Iodine group 3 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | Iodine group 4 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | Air cylinders | 4 | 4 | 4 | 4 | 4 | 4 | 4 | 4 | 4 | 4 | Aluminum cylinders | 4 | 4 | 4 | 4 | 4 | 4 | 4 | 4 | 4 | 4 |
| X-ray parameters | kVp | 106 | 90 | 102 | 90 | 102 | 90 | 102 | 103 | 102 | 103 | Tube current (mA) | 10 | 744 | 9 | 744 | 6.6 | 744 | 6.6 | 339 | 6.6 | 339 | Dose setting | Normal | Normal | Low | Normal | Low | Normal | Low | Normal | Low | Low | Frame rate | 7.5 | 15 | 7.5 | 15 | 5 | 15 | 5 | 10 | 5 | 10 | Added filteration | Al 2.0 + Cu 0.1 | Al 1.0 + Au 0.01 | Al 1.5 + Cu 0.6 | Al 1.0 + Au 0.01 | Al 1.5 + Cu 0.6 | Al 1.0 + Au 0.01 | Al 1.5 + Cu 0.6 | Al 1.0 + Au 0.01 | Al 1.5 + Cu 0.6 | Al 2.0 + Cu 0.1 |
| Temporal image quality | Moving wires | 2 | 3 | 1 | 3 | 1 | 3 | 1 | 3 | 1 | 3 |
|
|