Radiation Dose Optimization in Interventional Cardiology: A Teaching Hospital Experience

Table 6

Quantitative image quality results using the CIRS model 901 30 cm phantom for standard NEMA measurements.

Phase 1Phase 2Phase 3Phase 4Phase 5
FluoroscopyAcquisitionsFluoroscopyAcquisitionsFluoroscopyAcquisitionsFluoroscopyAcquisitionsFluoroscopyAcquisitions

X-ray parameterskVp118901139011090110102110102
Tube current (mA)11.4746117467.17467.13417.1340
Dose settingNormalNormalLowNormalLowNormalLowNormalLowLow
Frame rate7.5157.515515510510
Added filterationAl 2.0 + Cu 0.1Al 1.0 + Au 0.01Al 1.5 + Cu 0.6Al 1.0 + Au 0.01Al 1.5 + Cu 0.6Al 1.0 + Au 0.01Al 1.5 + Cu 0.6Al 1.0 + Au 0.01Al 1.5 + Cu 0.6Al 2.0 + Cu 0.1

Static image qualityLine pairs1.61.81.41.81.21.81.21.61.21.6
Iodine group 14646262626
Iodine group 20000000000
Iodine group 30000000000
Iodine group 40000000000
Air cylinders4444444444
Aluminum cylinders4444444444

X-ray parameterskVp106901029010290102103102103
Tube current (mA)1074497446.67446.63396.6339
Dose settingNormalNormalLowNormalLowNormalLowNormalLowLow
Frame rate7.5157.515515510510
Added filterationAl 2.0 + Cu 0.1Al 1.0 + Au 0.01Al 1.5 + Cu 0.6Al 1.0 + Au 0.01Al 1.5 + Cu 0.6Al 1.0 + Au 0.01Al 1.5 + Cu 0.6Al 1.0 + Au 0.01Al 1.5 + Cu 0.6Al 2.0 + Cu 0.1

Temporal image qualityMoving wires2313131313