Applicability of a Novel Wearable Wireless Electrocardiogram Monitoring Device (Spyder) for Arrhythmia Detection in Patients with Suspected Cardiac Arrhythmias
Table 4
Timing of arrhythmia detection.
Indications
Detection in the first 24 h
Detection in the first 48 h
Time to first detection (h)
Tachycardia
No
Yes
47
Ventricular fibrillation
No
No
No
Sinus arrest >3 s
No
No
No
Supraventricular tachycardia
No
Yes
47
Degree II A-V block
No
No
No
Degree III A-V block
No
No
74.4
Atrial fibrillation
No
Yes
28.8
Atrial flutter
No
No
No
Premature ventricular contraction
Yes
Yes
7.2