Safety and Feasibility Using a Fluid-Filled Wire to Avoid Hydrostatic Errors in Physiological Intracoronary Measurements
Table 1
Summary of simultaneous measurements including height measurements (n = 19).
| Index | Sensor-tipped wire | Fluid-filled wire | Difference between sensor-tipped and fluid-filled wire | Difference between sensor-tipped and fluid-filled wire after correction of the sensor-tipped wire for the hydrostatic effect |
| Pd/Pa | Mean ± SD | 0.93 ± 0.041 | 0.95 ± 0.035 | 0.024 ± 0.028 | 0.002 ± 0.018 | Median | 0.93 | 0.97 | 0.036 | 0.003 | Min | 0.84 | 0.88 | −0.04 | −0.06 | Max | 1.01 | 1.00 | 0.05 | 0.03 |
| FFR | Mean ± SD | 0.85 ± 0.066 | 0.87 ± 0.072 | 0.026 ± 0.033 | 0.0037 ± 0.017 | Median | 0.85 | 0.91 | 0.032 | 0.0016 | Min | 0.74 | 0.73 | −0.035 | −0.027 | Max | 0.96 | 0.96 | 0.072 | 0.03 |
|
|