Safety and Feasibility Using a Fluid-Filled Wire to Avoid Hydrostatic Errors in Physiological Intracoronary Measurements

Table 1

Summary of simultaneous measurements including height measurements (n = 19).

IndexSensor-tipped wireFluid-filled wireDifference between sensor-tipped and fluid-filled wireDifference between sensor-tipped and fluid-filled wire after correction of the sensor-tipped wire for the hydrostatic effect

Pd/PaMean ± SD0.93 ± 0.0410.95 ± 0.0350.024 ± 0.0280.002 ± 0.018
Median0.930.970.0360.003
Min0.840.88−0.04−0.06
Max1.011.000.050.03

FFRMean ± SD0.85 ± 0.0660.87 ± 0.0720.026 ± 0.0330.0037 ± 0.017
Median0.850.910.0320.0016
Min0.740.73−0.035−0.027
Max0.960.960.0720.03